x-rite i1 pro 2 with case and accessories.The Item is used like new,
SOLD AS IS, THE DEVICE HAS BEEN TESTED AND IS ON GOOD WORKING CONDITIONS
, No license included.
All sales are final, Returns no accepted.
PLEASE READ THE BELOW TEST RESULT BEFORE COMPLETING THE PURCHASE TO VOID MISS UNDERSTANDING'S
i1Pro Tests
General Information
Date and Time: 8/22/2021 4:06:57 PM
Application version: 4.2.5.381, 32-bit
SDK version i1Pro: 4.2.4.5061
SDK version i1iO: 2.2.4.5061
SDK version i1Pro3: 1.0.15.5919
SDK version i1iO3: 1.0.15.5919
SDK version i1iSis: 2.0.0.4968
SDK version ColorMunki: 1.0.3.4630
SDK version i1Display2: X-Rite, Inc. SipCal Library Ver. 7.3.7 (32 bit)
Platform: Microsoft Windows 10, 64-bit
Open instrument
Connection: Successful
Open instrument: Successful
Instrument information
Instrument inspection: Successful
Serial number i1Pro: 1081229
Type: i1Pro
Revision: E
Firmware version: 634
CPLD version: 999
Dark measurement count: 234
Lamp burning time: 787.2 seconds
Spot measurement count: 82
Scan measurement count: 118
Factory calibration: 4/11/2017
Spectral Module ID: 01-85b15e190000b7
Supplier: 'EFI-i1Pro'
Recertification date: -
Instrument Licenses
Measurement Modes: [Reflectance Spot, Reflectance Scan, Emission, DualReflectance Spot, DualReflectance Scan]
Licensed functionality: [Has UV LED, Has Wavelength LED, Has Zebra Ruler sensor, Max Ruler Length = 300 mm, Has Indicator LED, Spectral values, Tristimulus, Density]
Licensed Modules:
i1Match Module "Scanner": NOT licensed
i1Match Module "Monitor": NOT licensed
i1Match Module "Printer": NOT licensed
i1Match Module "Printer RGB only": NOT licensed
i1Match Module "Printer CMYK only": NOT licensed
i1Match Module "Beamer": NOT licensed
i1Match Module "Printer Easy RGB": NOT licensed
i1Match Module "Printer Easy CMYK": NOT licensed
i1Match Module "Digicam": NOT licensed
i1Match Module "Editor": NOT licensed
i1Profiler Module "Display Profiling": NOT licensed
i1Profiler Module "RGB Printer Profiling": NOT licensed
i1Profiler Module "CMYK Printer Profiling": NOT licensed
i1Profiler Module "Printer QA": NOT licensed
i1Profiler Module "Advanced-Mode": NOT licensed
i1Profiler Module "Projector": NOT licensed
Button test
Button: Successful
Emission test
Emission calibration: Successful
Prepare the instrument for emission measurement
White measurement: Successful
Gray measurement: Successful
Black measurement: Successful
Red measurement: Successful
Green measurement: Successful
Blue measurement: Successful
Emission: Successful
Button test
Button: Successful
Noise Tests
Noise measurements: Successful
Noise result: Successful
Reflectance test
Reflectance Spot calibration: Successful
Reflectance Spot measurement 1: Successful
Reflectance Spot measurement 2: Successful
Reflectance Spot measurement 3: Successful
Reflectance Spot measurement 4: Successful
Reflectance Spot measurement 5: Successful
Reflectance Spot measurement 6: Successful
Reflectance Spot measurement 7: Successful
Reflectance Spot measurement 8: Successful
Reflectance Spot measurement 9: Successful
Reflectance Spot measurement 10: Successful
Lamp test: Successful
Reflectance drift test: Successful
Reflectance Scan calibration: Successful
Dual Reflectance Spot calibration: Successful
Dual Reflectance Spot test: Successful
Dual Reflectance Scan calibration: Successful
Positioning Sensor Test
Scan with positioning sensors: Successful (Moving distance: 26.3 cm)
*** PASS ***
**************************
i1Pro Tests
General Information
Date and Time: 2021-08-22T16:06:57
Application version: 4.2.5.381, 32-bit
SDK version i1Pro: 4.2.4.5061
SDK version i1iO: 2.2.4.5061
SDK version i1Pro3: 1.0.15.5919
SDK version i1iO3: 1.0.15.5919
SDK version i1iSis: 2.0.0.4968
SDK version ColorMunki: 1.0.3.4630
SDK version i1Display2: X-Rite, Inc. SipCal Library Ver. 7.3.7 (32 bit)
Platform: Microsoft Windows 10 (Build 19041), 64-bit
Open instrument
Connection: Successful
Open instrument '\?\usb#vid_0971&pid_2000#5&26f3984c&0&2#{989d6515-e122-4442-a357-751457f102a5}': Successful
Instrument information
Instrument inspection: Successful
Serial number i1Pro: 1081229
Type: i1Pro
Revision: E
Firmware version: 634
CPLD version: 999
Dark measurement count: 234
Lamp burning time: 787.2 seconds
Spot measurement count: 82
Scan measurement count: 118
Factory calibration: 2017-04-11
Spectral Module ID: 01-85b15e190000b7
Supplier: 'EFI-i1Pro'
Recertification date: -
Instrument Licences
Measurement Modes: [ReflectanceSpot, ReflectanceScan, EmissionSpot, DualReflectanceSpot, DualReflectanceScan]
Licensed functionality: [Has UV LED, Has Wavelength LED, Has Zebra Ruler sensor, Max Ruler Length = 300 mm, Has Indicator LED, Spectral values, Tristimulus, Density]
Licensed Modules:
i1Match Module "Scanner": NOT licensed
i1Match Module "Monitor": NOT licensed
i1Match Module "Printer": NOT licensed
i1Match Module "Printer RGB only": NOT licensed
i1Match Module "Printer CMYK only": NOT licensed
i1Match Module "Beamer": NOT licensed
i1Match Module "Printer Easy RGB": NOT licensed
i1Match Module "Printer Easy CMYK": NOT licensed
i1Match Module "Digicam": NOT licensed
i1Match Module "Editor": NOT licensed
i1Profiler Module "Display Profiling": NOT licensed
i1Profiler Module "RGB Printer Profiling": NOT licensed
i1Profiler Module "CMYK Printer Profiling": NOT licensed
i1Profiler Module "Printer QA": NOT licensed
i1Profiler Module "Advanced-Mode": NOT licensed
i1Profiler Module "Projector": NOT licensed
Button Test
Button: Successful
Emission test
Emission calibration: Successful
Prepare the instrument for emission measurement
White measurement: Successful (X=388.474 Y=409.130 Z=511.597)
Gray measurement: Successful (X=137.767 Y=144.253 Z=181.045)
Black measurement: Successful (X=0.422 Y=0.433 Z=0.841)
Red measurement: Successful (X=165.999 Y=83.555 Z=13.127)
Green measurement: Successful (X=147.247 Y=304.539 Z=52.929)
Blue measurement: Successful (X=88.076 Y=34.195 Z=461.752)
Emission: Successful
Button Test
Button: Successful
Noise Tests
Noise measurements: Successful
Noise result: Successful (electronic noise: 3.430)
Reflectance test
Reflectance Spot calibration: Successful (slope=-0.000355809345 (0.150000006), WLS=-0.400000006)
Reflectance Spot measurement 1: Successful (95.801, -0.085, 2.528)
Reflectance Spot measurement 2: Successful (95.789, -0.074, 2.540)
Reflectance Spot measurement 3: Successful (95.781, -0.075, 2.544)
Reflectance Spot measurement 4: Successful (95.773, -0.068, 2.549)
Reflectance Spot measurement 5: Successful (95.773, -0.068, 2.545)
Reflectance Spot measurement 6: Successful (95.767, -0.068, 2.548)
Reflectance Spot measurement 7: Successful (95.767, -0.077, 2.566)
Reflectance Spot measurement 8: Successful (95.766, -0.072, 2.559)
Reflectance Spot measurement 9: Successful (95.771, -0.085, 2.577)
Reflectance Spot measurement 10: Successful (95.770, -0.092, 2.583)
Lamp test: Successful (minL=95.766 meanL=95.776 MIT=0.0143640004 CIT=0.0143640004 LOT=790.904785)
Reflectance drift test: Successful (0.019)
Reflectance Scan calibration: Successful (slope=-0.000355809345 (0.150000006), WLS=-0.349999994)
Dual Reflectance Spot calibration: Successful (slope=-0.000355809345 (0.150000006), WLS=-0.349999994)
Dual Reflectance Spot test: Successful
Dual Reflectance Scan calibration: Successful (WLS=-0.349999994)
Positioning Sensor Test
Scan with positioning sensors: Successful (min: -0.001, max: 0.262, total distance: 0.263)
*** PASS ***